Taper Delivery Mode
Ta
pe
r
Pr
og
ra
mmi
ng
9. Review the Plateau Rate
• The plateau rate is a calculated value and cannot be edited.
10. Enter the KVO Rate
• Press
U
or
D
to highlight
KVO Rate.
If the value shown is correct,
select
Accept Value
and go to step 11. If the value is not correct, press
S
.
• Press
U
or
D
until the value reads
0 mL/hr
and select
Save.
• The pump then returns to the Review screen.
11. Enter the Reservoir Volume
• Press
U
or
D
to highlight
Reservoir Vol.
If the value shown is
correct, select
Accept Value
and go to step 12. If the value is not
correct, press
S
.
• Press
U
or
D
until the value reads
1050 mL
and select
Save.
• The pump then returns to the Review screen.