118
References and Troubleshooting
Re
fe
re
nc
es a
nd
Tr
ou
bl
es
ho
ot
ing
Alarm / Message
Alarm
Priority
Description / Corrective Action
Unusable battery.
Pump stopped.
High
The batteries installed are either the wrong type of AA batteries,
or the rechargeable battery pack is not compatible with the pump.
Select
Acknowledge
to clear the alarm.
Remove the battery or batteries from service. Install 4 new AA
batteries or a fully charged rechargeable battery pack. To start
delivery, good batteries must always be installed even when an
external source of power is connected. If appropriate, start the
pump
Unusable battery.
Pump will not run.
Medium The batteries installed are either the wrong type of AA batteries,
or the rechargeable battery pack is not compatible with the pump.
Select
Acknowledge
to clear the alarm.
Remove the battery or batteries from service. Install 4 new AA
batteries or a fully charged rechargeable battery pack. To start
delivery, good batteries must always be installed even when an
external source of power is connected. If appropriate, start the
pump.
Upstream occlusion.
Clear occlusion
between pump and
reservoir.
High
Fluid is not flowing from the fluid container to the pump, which
may be due to a kink, closed clamp, or air bubble in the tubing
between the fluid container and pump. Delivery is paused and
will resume if the occlusion is removed. Remove the obstruction
to resume operation. The alarm will clear when the occlusion is
removed. You must acknowledge this alarm after it clears if it has
occurred and cleared more than 3 times within 15 minutes.
NOTE:
This alarm has associated help screens.