26
Pump Programming
Pu
mp
Pr
og
ra
mmi
ng
Pump Programming
Protocols and Protocol Libraries
A
protocol
is a collection of medication delivery and other pump settings, including administrator
settings for the CADD
®
-Solis VIP ambulatory infusion pump, a therapy, qualifier, and drug combination,
and other delivery parameters. A collection of protocols in CADD
®
-Solis Medication Safety Software form
a
protocol library.
The CADD
®
-Solis system administrator creates, maintains, and loads the libraries onto
the CADD
®
-Solis VIP ambulatory infusion pump.
Each protocol includes a therapy, qualifier, and drug.
•
Therapies
are pre-defined and correspond to the delivery modes available on the pump: PCA,
continuous, intermittent, step, and taper. No additional therapies can be created.
•
Qualifiers
are specific to a selected therapy and further identify the type of infusion to be delivered.
Examples of qualifiers include pediatric, TPN, and epidural.
•
Drugs
are designated by a name/concentration/units/drug ID combination.
New Protocol Selection
Single Protocol Download
1. Make sure the pump is stopped and on the home screen.
2. Connect the pump to the PC using the USB cable. Use the CADD
®
-Solis Medication Safety Software to
load a single protocol on the pump.
3. Acknowledge the low priority alarm.
4. The “Review pump settings” screen appears. Select
Review.
5. Press
U
or
D
to scroll through and review all the settings. If the
highlighted setting is correct, press
Accept Value.
A green check mark
appears next to the setting.
If the highlighted setting is incorrect, press
S
to adjust the setting.
Press
U
or
D
to modify the setting and then press
Save.
6. When your review is complete, select
Next.
7. If a cassette is attached, latched, and locked (if required), the “Start
pump?” screen appears. Press
Yes
to start the pump immediately or
No
to start the pump at a later time.