84
Operating the Pump
O
pe
ra
ti
ng
the P
ump
Prime Tubing, No Cassette Change
1. Stop the pump if it is running.
2. In the Tasks menu (see page 88), press
U
or
D
until
Prime Tubing
is
highlighted, and then press
S
.
3. Unlock the keypad, if required.
4. If you have not already done so, disconnect the tubing from the patient,
open the clamps, and select
Prime.
WARNING:
Do not prime the fluid path with the tubing connected
to a patient as this could result in overdelivery of medication
or air embolism, which could result in serious patient injury or
death.
5. Stop priming at any time by selecting
Stop Priming.
Priming
automatically stops after 10 mL (or 20 mL if a high volume set is
attached) are primed. Continue priming as needed.
WARNING:
To prevent air embolism, ensure that the entire fluid
path is free of all air bubbles before connecting to the patient.
Air embolism could result in serious patient injury or death.