Air and Occlusion Settings
In the Advanced Tasks menu, press
U
or
D
to highlight
Air and
Occlusions Settings
and press
S
.
Air Detector On/Off
The pump is built with an integrated air detector. The pump may be customized to use the air detector. If
air is detected in the part of the tubing that passes through the air detector, an alarm sounds and delivery
stops. If an air detector is not required, it may be turned off.
The Air Detector On/Off setting allows you to control whether the air detector is turned on or off. If the
air detector is turned on, an alarm sounds when air is detected in the fluid path. For certain therapies, for
example, subcutaneous infusions, it may be desirable to turn the air detector off.
WARNING:
If the air detector is turned off, the pump does not detect air in the fluid path. It is
recommended that you periodically inspect the fluid path and remove any air to prevent air
embolism, which could result in serious patient injury or death.
To turn the air detector on or off:
1. Make certain the pump is stopped.
2. In the Air and Occlusions Settings menu, press
U
or
D
to highlight
Air Detector On/Off
and press
S
.
3. Unlock the keypad.
4. Press
U
or
D
to highlight
On
or
Off,
and select
Save.