Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
The CADD®-Solis pump is intended for use in the electromagnetic environment specified below. The customer or user of the
CADD®-Solis pump should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment—Guidance
Portable and mobile RF communications equipment should
be used no closer to any part of the Pump, including cables,
than the recommended separation distance calculated from
the equation applicable to the frequency of the transmitter.
Conducted RF IEC
61000-4-6
3 Vrms
150 kHz to 80 MHz
outside ISM bandsa
3 Vrms
Recommended separation distance
d = 1.2
Conducted RF IEC
61000-4-6
10 Vrms
150 kHz to 80 MHz
in ISM bandsb
10 Vrms
Recommended separation distance
d = 1.2
Radiated RF IEC
61000-4-3
10 V/m
80 MHz to 2.5 GHz
10 V/m
d = 1.2
80 MHz to 800 MHz
d = 2.3
800 MHz to 2.5 GHz
Where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d
is the recommended separation
distance in meters (m).c
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey,
c
should be less than the
compliance level in each frequency range.d
Interference may occur in the vicinity of equipment marked
with the following symbol:
R
NOTE 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people
a The ISM (industrial, scientific, and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to
13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
b The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5
GHz are intended to decrease the likelihood that mobile/portable communications equipment could cause interference if it is
inadvertently brought into patient areas. For this reason, an additional factor of 10/3 is used in calculating the recommended
separation distance for transmitters in these frequency ranges.
c Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the CADD®-Solis pump is used exceeds the applicable RF compliance
level above, the CADD®-Solis pump should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the CADD®-Solis pump.
d Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.