Autolock
The CADD
®
-Solis VIP ambulatory infusion pump is designed to meet both safety and usability needs. The
autolock feature reduces the chance of unauthorized pump programming. When the keypad is unlocked
with a security code and left unlocked, the software automatically locks the keypad.
If the pump is on the home screen, the autolock feature takes effect 30 seconds after the last key press.
Autolock takes longer on programming or task screens when the user typically needs more time to
perform an action. Depending on which screen was on the pump last, and if the pump is not alarming, it
can take up to 4 minutes after the last key press before the pump reverts to the home screen and autolocks
immediately. When the pump is alarming, autolock does not take effect.
NOTE:
The keypad can be re-locked by pushing the right soft key twice from the home screen or once
from the Tasks or Advanced Tasks menu. As a recommended safety precaution, always manually lock
the pump using this feature.
Entering Security Codes
WARNING:
•
Do not disclose pump security codes or any other information that would allow the patient or
unauthorized clinician complete access to programming and operating functions. Improper
programming could result in serious patient injury or death.
•
Do not leave the pump unattended while unlocked. All programming functions are accessible while
the pump is unlocked, and improper programming could result in serious patient injury or death.
NOTE:
Many of the instructions in this manual include steps to unlock the keypad. This step is required
only if the keypad is locked.
To enter a security code when prompted:
1. Press
U
or
D
to scroll up or down to the correct digit. Press
S
or
Accept Value
to advance to the next digit. Once the complete code
has been entered, press
S
or
Accept Value.
2. If an incorrect code is entered, a wrong code error appears. It includes
the security level required to access the function. Select
Retry
to enter
the code again.
NOTE:
If you enter a code that you believe is correct and receive
a wrong code message, check the screen to see if the code has
been customized. If the code is not customized, it was entered
incorrectly and you should retry. If the code is customized, enter
the customized code. If you do not know the code, contact your
CADD
®
-Solis system administrator.