Intermittent Delivery Mode
In
te
rmi
tte
nt
Pr
og
ra
mmi
ng
KVO Rate
The KVO or “keep vein open” rate is optional. It allows the delivery of a
minimal amount of drug to help maintain catheter patency.
If a next dose start time is programmed, the KVO rate is active until the
dose begins. It is also active between doses.
Reservoir Vol.
This is the volume of fluid contained in the reservoir. The reservoir volume
value decreases as the pump is primed or delivers fluid.
The CADD
®
-Solis system administrator may choose a standard reservoir
volume which allows the reservoir volume to be quickly reset to that value.
The clinician can edit the reservoir volume to another amount. This change
is for the current protocol only and remains in effect until the Start New
Patient or Same Patient New Protocol task is selected.
Programming Example
WARNING:
Ensure that the ± 6% system delivery accuracy specification is taken into account when
programming the pump and/or filling the reservoir. Failure to do so may result in medication in
the reservoir becoming depleted sooner than expected. If the pump is used to deliver critical or life-
sustaining medication, the interruption in the delivery of medication could result in patient injury
or death.
Example:
Medication is provided in a 100 mL CADD™ medication cassette reservoir. The patient must
receive a 23.5 mL dose over 1 hour. The dose must be given every 6 hours with a 0.2 mL/hr KVO rate
between doses. The first dose must begin at 8:00 pm tonight.
1. Select the Therapy
• Stop the pump if it is running.
• Select
Tasks,
then
View Advanced Tasks,
and then
S
tart New
Patient.
• Unlock the keypad.
• Press
U
or
D
to highlight the
Intermittent
therapy and press
S
. See
on page 27 if you need more information
on selecting the therapy.
2. Select the Qualifier
• Press
U
or
D
to highlight the desired qualifier and press
S
.