Numerical Calculations
19-14
19
Press the SET key, the cursor will move to the right. Press the SET key again and a popup
menu will appear. Select the "-" and press the SET key.
- The cursor will appear to the right of the "-." Press the SET key and a popup menu will appear.
Select a "TYPE" and press the SET key.
00X0 will be displayed in the formula field.
00X0 refers to the X coordinate of the detection point in model 0, stored in register NO.00 (image
0).
3. Move the cursor to the 2nd line using the up and down arrow keys and set the upper limit value
using the left and right arrow keys. Then press the SET key. Then select a digit to change using
the left and right arrow keys. Repeat this procedure to set the value to +160.0000 and then press
the SET key.
Move the cursor to the lower limit using the left and right keys, and press the SET key. Select a
digit using the left and right keys Enter the 140.0000 using the up and down keys, and
press the SET key.
+0000140.0000 to +0000160.0000 will be displayed in the lower and upper limit fields.
4. Move the cursor to the "N01" line and the "TYPE" column using the up/down keys.
- As described in steps 2 to 3, enter 01Y0 to 00Y0 in the formula field, and inter +0000090.0000
to +0000095.0000 in the upper and lower limit fields.
5. Move the cursor to the "
RUN A TEST" item using the up and down keys, and press the SET
key.
Press the SET key once more. Then the settings will be stored, and the test will be executed.
The evaluation result (OK/NG) of the calculations will be displayed.
6. Press the ESC key to return to the "MEA-CND" setting screen.
N00
X
01X0−00X0
+
0000000.00〜
+
0000000.00
+0000000.0000~+0000160.0000
Upper limit
+0000140.0000~+0000160.0000
Lower limit
N00
X
01X0−00X0
+
0000140.0000〜
+
0000160.0000
N01
Y
01Y0−00Y0
+
0000090.0000〜
+
0000095.0000
N00
X
01X0−00X0
+
0000140.0000〜
+
0000160.0000
+
0000147.0000 OK
N01
Y
01Y0−00Y0
+
0000090.0000〜
+
0000095.0000
+
0000091.0000 OK
OK: When the test result based on the calculations is within the specified range
NG: When the test result based on the calculations is outside of the specified range