Before start-up check whether the rated voltage of the device coincides
with the local line voltage .
The device was developed for laboratory use and must be protected from
every type of fluid or vapour .
The device can be operated b5°C and +40°C .
The welding parameters are qualified for room temperature (20°C – 25°C)
operation only .
Optimal relative humidity is 60% .
The device is programmed with standard parameter sets for welding
C-Flex
®
374, AdvantaFlex
®
, PharMed
®
BPT and SaniPure
™
BDF
™
.
These parameter sets have been validated at ambiant temperature .
The degree of protection is IP20 .
Make sure that the device is not switched off during the welding process .
Ensure that there is an uninterrupted power supply (UPS) .
Use only the blades provided by the manufacturer and use it only once .
Use only tested control cables and accessories .
Isolate the supply voltage before replacing a fuse .
The device has an Ethernet connection . Only skilled personnel may
connect the device to a network .
Only shielded Ethernet cables may be used .
Only skilled personnel may configure the device for connection to a
network .
It is recommended to operate the device via the touch screen during
sequences .
The device contains electric drives which can be blocked | destroyed by
foreign substances . Make sure that no foreign substances reach the
interior of the device .
After using the Biowelder
®
TC, you should always log out . Otherwise
another user is able to manipulate the device with your account .
The auto login is used to work more efficiently with the Biowelder
®
TC .
The auto login can only be activated for one user . This user will be logged in
AUTOMATICALLY without entering any username and password after
each restart or after changing the user interface .
A user of the Supervisor group should NOT use the auto login because
of safety reasons, otherwise any user is able to take essential changes on
the Biowelder
®
TC under the Supervisor account .
Not following the manual instructions of the manufacturer may lead in to
unreliability, hazardous situations and disturbances of safety .
Содержание Biowelder Total Containment
Страница 2: ...English page 3 Deutsch Seite 47 Fran ais page 93 Espa ol p gina 139 Italiano pagina 185...
Страница 92: ...92 Bedienungsanleitung Biowelder Total Containment...
Страница 138: ...138 Mode d emploi Biowelder Total Containment...
Страница 184: ...184 Manual de funcionamiento Biowelder Total Containment...
Страница 230: ...230 Manuale d uso Biowelder Total Containment...
Страница 231: ...Manuale d uso Biowelder Total Containment 231...