4-6
Error number
Message
Cause
Action
230
ABNORMAL CHAMBER 2
SENSOR READING: LOW
TEMP.
Disconnected temperature sensor Contact the Sakura authorized distributor.
231
ABNORMAL CHAMBER 2
SENSOR READING: HIGH
TEMP.
Short-circuited temperature sensor
232
TIME OUT: CHAMBER
HEATING
Temperature did not rise within a
specified time.
240
OVERPRESSURE:
COMPRESSOR
The refrigerator error was found.
250
BRINE PUMP STOPPED
The brine pump stopped.
300, 301
DRAIN VALVE FAILURE
An error occurred opening/closing
the drain valve.
302 to 309
VACUUM CIRCUIT ERROR
An error occurred in the vacuum
circuit controlling the
opening/closing of the valve.
310 to 312
VACUUM ERROR
An error occurred controlling the
chamber negative
pressure/negative pressure
cancellation.
If the window is open, close the window. Clean
the gasket contacting the window, if dirty. If the
problem persists, contact the Sakura authorized
distributor.
320, 321, 331
PRESSURE SENSOR
ERROR
The pressure sensor status change
expected during vacuum circuit
operation does not occur.
Contact the Sakura authorized distributor.
330
OZONE PORT IS CLOGGED. The ozone port got clogged.
Clean the ozone port.
340
OZONE LAMP ERROR
The ozone sensor reacts although
the ozone lamp is not lit.
Contact the Sakura authorized distributor.
341
OZONE LAMP ERROR
The ozone sensor does not react
although the ozone lamp is lit.
Contact the Sakura authorized distributor.
350
WINDOW OPEN ERROR
The window is open at the start of
ozone cycle.
If the window is open, close the window. If the
problem persists, contact the Sakura authorized
distributor.
360
WINDOW LOCK ERROR
Unable to Lock the Window.
Open the window fully and then close the
window. If the problem persists, contact the
Sakura authorized distributor.
361
WINDOW UNLOCK ERROR
Unable to Unlock the Window.
Contact the Sakura authorized distributor.
370
OZONE SENSOR ERROR
Ozone was detected not during
ozone cycle.
Contact the Sakura authorized distributor.
371
OZONE SENSOR ERROR
Ozone was not detected after the
ozone lamp remained lit for a
specified period of time.
This error may occur when a lot of acetone or
alcohol is used. Wait for a while and perform
ozone cycle. If the problem persists, contact the
Sakura authorized distributor.
400, 401, 402
CHUCK HOME POSITION
ERROR
An error occurred searching for the
specimen head home.
Contact the Sakura authorized distributor.
410, 411
HANDWHEEL HOME
POSITION ERROR
An error occurred searching for the
handwheel home with the motor
running.
Motorized sectioning cannot be performed, but
manual sectioning can be performed. Check the
operating location for any obstruction. Perform
motorized sectioning again. If the problem is
repeated, contact the Sakura authorized
distributor.
420
CHUCK POSITION ERROR
Obstruction, such as shavings. in or
around the chuck head and/or
specimen holder
Check for obstruction and remove if any. If the
problem persists, contact the Sakura authorized
distributor.
421
HANDWHEEL LOCKED
WHILE IN MOTION
The handwheel lock tab was
engaged during motorized
sectioning.
Disengage the handwheel lock tab.
430
EMERGENCY STOP
An emergency stop was actuated
with the emergency stop button.
Release the emergency stop button.
440
STOPPED FAN
Fan operation error was detected. Contact the Sakura authorized distributor.
441
OPEN FUSE
A blown fuse was detected.
Содержание Tissue-Tek Polar DM
Страница 1: ...Operating Manual Frozen Tissue Section Preparation Instrument Cryostat Microtome...
Страница 15: ...Chapter 1 Basic Knowledge of Instrument...
Страница 32: ...1 17 Standard Accessories Options 4 5 15...
Страница 68: ...1 53...
Страница 69: ...Chapter 2 Operating Method...
Страница 108: ...2 39...
Страница 109: ...Chapter 3 Maintenance Inspection...
Страница 122: ...3 13...
Страница 123: ...Chapter 4 Troubleshooting...