4-1
Troubleshooting
Before Suspecting a Trouble ~ Instrument Conditions and Remedial Actions
The troubleshooting chart below lists problems that may occur during system operation. Possible causes
and remedial actions are described for each problem. Utilize this information to identify the system
condition and resolve the problem. If the problem cannot be resolved or the applicable condition is not listed,
contact the nearest authorized Sakura distributor or representative.
* If an unexpected problem occurred and the condition cannot be rectified easily, turn off the
power immediately. Contact the nearest authorized Sakura distributor or representative.
Symptom
Check
Action
Reference
Page
Nothing is displayed
on the screen.
Is the power switch turned on?
Turn on the power.
2-2
Isn't the power cable or plug disconnected?
Connect the power cable or plug.
1-4
Is the power supplied?
Check the breaker, etc., on the facility side.
2-2
The instrument power was temporarily cut off
(momentary power failure).
Turn off the instrument power and then turn it
back on.
Isn't the instrument in the sleep mode or standby
mode?
Touch the screen. The screen should come
back.
2-3
The draft generator
does not operate.
Isn't the connector unplugged?
Plug in the connector.
3-5
Is the draft generator power turned on?
Turn on the power.
2-35
Is the draft generator setting turned on?
Turn it on.
2-35
Are the LED lamps on the draft generator lit?
Check the lamp mode table on p. 3-5.
3-5
The vacuum suction
force dropped.
Aren't contaminants trapped in the vacuum
suction port?
Use tweezers, etc., to remove the
contaminants.
3-10
Aren't contaminants trapped on the vacuum
filter?
Replace the vacuum filter set.
3-4
Aren't contaminants trapped in the hose?
Remove the contaminants.
3-4
Isn't the connection between the vacuum filter
and hose loose?
Connect them securely.
3-3
Isn't the desired suction force set low?
Raise the set value of desired suction force.
1-36
The chamber cannot
be cooled.
Isn't the setpoint temperature high?
Readjust the setpoint temperature.
1-21 to 25
Is the operating environment temperature
appropriate?
Check the operating environment
temperature.
1-16
Want to perform
sectioning although
the chamber cannot
be cooled.
By using the emergency plastic holder
(optional), heat conduction from the specimen
head can be suppressed to make it difficult for
the specimen to melt.
2-37
Specimens cannot be
cut properly.
Aren't contaminants trapped on the blade
holder?
Clean the blade holder.
3-7, 8
Isn't the blade edge dirty?
Clean the blade edge or change the
disposable blade.
2-8
Hasn't the specimen melted due to insufficient
freezing of the block?
Freeze the block again.
2-38
Hasn't the sectioning position remained the
same?
The disposable blade has been consumed, so
change the disposable blade. Or, slide the
blade holder to change the position at which
the disposable blade contacts the specimen so
that the entire disposable blade can be used
evenly.
2-8, 9
Are the various levers (specimen head locking
lever, orientation locking lever, disposable blade
mounting lever, blade holder left/right lever, blade
holder mounting lever) locked?
Lock the levers.
2-7, 30
Are the blade holder coma and wave washer
installed correctly?
Install the blade holder coma and wave
washer correctly.
3-8
Are the chamber temperature and specimen
head temperature appropriate?
Adjust the setpoint temperatures.
1-21 to 25
Are the setpoint temperature and set thickness
appropriate?
Adjust the setpoint temperature and set
thickness.
1-21 to 25
The handwheel does
not move.
Isn't the handwheel lock tab engaged?
Disengage the handwheel lock tab.
1-8
Isn't motorized sectioning paused?
Press the stop key to cancel the pause.
1-11
Motorized sectioning
(right panel) cannot be
performed.
Select the MOTORIZED Tab.
Select the MOTORIZED Tab.
2-26
Isn't the handwheel lock tab engaged?
Disengage the handwheel lock tab.
1-8
Содержание Tissue-Tek Polar DM
Страница 1: ...Operating Manual Frozen Tissue Section Preparation Instrument Cryostat Microtome...
Страница 15: ...Chapter 1 Basic Knowledge of Instrument...
Страница 32: ...1 17 Standard Accessories Options 4 5 15...
Страница 68: ...1 53...
Страница 69: ...Chapter 2 Operating Method...
Страница 108: ...2 39...
Страница 109: ...Chapter 3 Maintenance Inspection...
Страница 122: ...3 13...
Страница 123: ...Chapter 4 Troubleshooting...