2-31
Other Operations
Adjusting the Blade Angle (Relief Angle)
You can adjust the blade angle by using an Allen wrench to
loosen counterclockwise (tilting the Allen wrench away from you)
the blade angle adjustment bolt found on the left side face of the
blade holder.
The recommended angle (relief angle) stamped on the blade
holder is 8 to 10
°
.
Once the position is fixed, use the Allen wrench to tighten the bolt
clockwise (tilt the Allen wrench away from you) and check if the
holder moves.
Blade angle
adjustment
bolt (hex)
Tighten