![Sakura Tissue-Tek Polar DM Скачать руководство пользователя страница 104](http://html1.mh-extra.com/html/sakura/tissue-tek-polar-dm/tissue-tek-polar-dm_operating-manual_3543386104.webp)
2-35
Other Operations
Draft Generator
The draft generator provides the function to suction air in the top part of the chamber from both sides when the window
is opened, in order to keep contaminated air from diffusing out of the chamber. Suctioned air is passed through the
HEPA filter in the rear of the instrument and then exhausted to the outside.
The draft generator is optional.
The draft generator can be turned on/off using the "Draft Generator" button on the [UTILITIES] Tab Screen.
Starting the Draft Generator
1. Press the "Draft Generator" button on the [UTILITIES] Tab
Screen.
2. The draft generator setting screen appears.
3. Press the "ON" button.
4. Pressing the “Enter” button saves the setting, after which the
display returns to the [UTILITIES] Tab Screen.
The draft generator starts operating when the window is opened.
If no operation has been performed for 5 minutes, the setting
is not saved and the screen closes automatically.
Stopping the Draft Generator
1. Press the "Draft Generator" button on the UTILITIES tab
screen.
2. The draft generator setting screen appears.
3. Press the "OFF" button.
4. Pressing the “Enter” button saves the setting, after which the
display returns to the [UTILITIES] Tab Screen.
The draft generator stops operating.
If no operation has been performed for 5 minutes, the setting
is not saved and the screen closes automatically.
Changing the Suction Force of the Draft Generator
To change the suction force of the draft generator, press the
"Suction Force Change" switch on the instrument.
The suction level can be switched over 5 levels (1 to 5), with the
suction force increasing by one level every time the "Suction Force
Change" switch is pressed. After reaching the highest level of 5,
the suction force returns to 1 the next time the switch is pressed.
The recommended suction level is "5."
LED lamp
LED lamp
Draft generator
Power switch
Suction force
change switch
Draft Generator button
ON button
OFF button
Enter button
Содержание Tissue-Tek Polar DM
Страница 1: ...Operating Manual Frozen Tissue Section Preparation Instrument Cryostat Microtome...
Страница 15: ...Chapter 1 Basic Knowledge of Instrument...
Страница 32: ...1 17 Standard Accessories Options 4 5 15...
Страница 68: ...1 53...
Страница 69: ...Chapter 2 Operating Method...
Страница 108: ...2 39...
Страница 109: ...Chapter 3 Maintenance Inspection...
Страница 122: ...3 13...
Страница 123: ...Chapter 4 Troubleshooting...