Remote Control
R&S
®
RTM20xx
295
User Manual 1317.4726.02 ─ 01
Table 15-13: Header data
Position
Meaning
Example
1
XStart in s
-9.477E-008 = - 94,77 ns
2
XStop in s
9.477E-008 = 94,77 ns
3
Number of samples
200000
4
Number of values per sample interval. For envelope
waveforms the value is 2.
2
Suffix:
<m>
.
Return values:
<DataHeader>
Comma-separated value list
Example:
-9.477E-008,9.477E-008,200000,2
Usage:
Query only
CHANnel<m>:DATA:POINts
<Points>
As a setting, the command selects a range of samples that will be returned with
number of returned samples for the selected range.
If
MSAMples
(maximum sample rate), the memory usually
contains more data samples than the screen can display. In this case, you can decide
which data will be saved: samples stored in the memory or only the displayed samples.
Note:
The sample range can be change only in STOP mode. If the acquisition is running,
DEF is always used automatically. If the acquisition has been stopped, data can be read
from the memory, and all settings are available.
Suffix:
<m>
.
The command affects all channels, and the suffix is irrelevant.
Command Reference