Remote Control
R&S
®
RTM20xx
369
User Manual 1317.4726.02 ─ 01
SEARch:TRIGger:DATatoclock:CLEVel
<ClockLevel>
Sets the voltage level for the clock signal. Clock level and clock edge define the reference
point for setup and hold time.
Parameters:
<ClockLevel>
Range:
depends on vertical scale
Firmware/Software:
FW 03.800
SEARch:TRIGger:DATatoclock:DLEVel
<DataLevel>
Sets the voltage level for the data signal. The data lavel defines the point of data transition.
Parameters:
<DataLevel>
Range:
depends on vertical scale
Firmware/Software:
FW 03.800
SEARch:TRIGger:DATatoclock:CLEVel:DELTa
<LevelDelta>
SEARch:TRIGger:DATatoclock:DLEVel:DELTa
<LevelDelta>
Set a hysteresis range to the clock and data levels in order to avoid unwanted search
results caused by noise oscillation around the level. For a rising edge, the hysteresis is
below the search level. Otherwise, for a falling edge the hysteresis is above the level.
Parameters:
<LevelDelta>
Range:
Lower limit depends on vertical scale and other set-
tings, no upper limit
Firmware/Software:
FW 03.800
SEARch:TRIGger:DATatoclock:CEDGe
<ClockEdge>
Sets the edge of the clock signal to define the time reference point for the setup and hold
time.
Parameters:
<ClockEdge>
POSitive | NEGative | EITHer
*RST:
POS
Firmware/Software:
FW 03.800
SEARch:TRIGger:DATatoclock:HTIMe
<HoldTime>
Sets the minimum time
after
the clock edge while the data signal must stay steady above
or below the data level. The hold time can be negative. In this case, the hold time ends
before the clock edge, and the setup time must be positive and longer than the absolute
value of the hold time.
Parameters:
<HoldTime>
Range:
depends on time base and sample interval
Command Reference