Remote Control
R&S
®
RTM20xx
371
User Manual 1317.4726.02 ─ 01
Parameters:
<Function>
AND | OR | NAND | NOR
AND
The required states of all channels must appear in the input signal
at the same time.
OR
At least one of the channels must have the required state.
NAND
"Not and" operator, at least one of the channels does not have the
required state.
NOR
"Not or" operator, none of the channels has the required state.
*RST:
AND
SEARch:TRIGger:PATTern:LEVel<n>
<ThresholdLevel>
Sets the threshold value for each specified source channel. You can set different levels
for the channels
Suffix:
<n>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<ThresholdLevel>
Range:
Depends on vertical scale
SEARch:TRIGger:PATTern:LEVel<n>:DELTa
<LevelDelta>
Sets a hysteresis range to the level of the specified source channel in order to avoid
unwanted search results caused by noise oscillation around the level. For a rising edge,
the hysteresis is below the search level. Otherwise, for a falling edge the hysteresis is
above the level.
Suffix:
<n>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<LevelDelta>
Range:
Lower limit depends on vertical scale and other set-
tings, no upper limit
SEARch:TRIGger:PATTern:WIDTh:RANGe
<Range>
Sets the condition how the duration of a steady pattern is compared with the given ref-
erence time.
To set the reference value
width
Command Reference