Masks
R&S
®
RTM20xx
126
User Manual 1317.4726.02 ─ 01
The specified factor in divisions is added to the y-values of the upper mask limit and
subtracted from the y-values of the lower mask limit. Thus, the upper half of the mask
is pulled upwards, the lower half is pulled down.
7. Similarly, to change the width of the waveform in horizontal direction, press "Width
X" and turn the "Navigation" rotary knob.
The left half of the mask is pulled to the left, the right half is pulled to the right.
8. To move the mask vertically on the screen, press "Y-Position" and turn the "Naviga-
tion" rotary knob to change the y-position.
The current position is indicated as an offset from the center in divisions; a positive
value indicates the waveform was moved upwards; a negative value indicates the
waveform was moved down.
9. To change the scaling of the mask in y-direction, press "Stretch Y" and turn the
"Navigation" rotary knob.
Turn the knob clockwise to increase the factor, or counterclockwise to decrease the
factor. A value over 100% stretches the amplitudes; a value less than 100% com-
presses the amplitudes.
10. Save the new mask:
a) Press "Save".
Working with Masks