Protocol Analysis
R&S
®
RTM20xx
204
User Manual 1317.4726.02 ─ 01
Fig. 11-16: Decoded LIN signal with frame table and applied label list
gray
= synchronization break, synchronization byte, correct checksum
yellow = identifier
green = parity bits
blue
= data words (UART words)
Table 11-5: Content of the LIN frame table
Column
Description
Start time
Time of frame start in relation to the trigger point
ID
Identifier value, hexadecimal value
Label
Symbolic label, available if a label list was loaded and applied
Length
Number of data bytes
Data
Hexadecimal values of the data bytes
Chks
Checksum value
State
Overall state of the frame.
Remote commands:
●
LIN (Option R&S RTM-K3)