Remote Control
R&S
®
RTM20xx
438
User Manual 1317.4726.02 ─ 01
Parameters:
<BusWidth>
Maximum number is the number of input channels.
Range:
1 to 16
Increment: 1
*RST:
4
Default unit: Bit
Parallel Clocked Bus - Configuration
.............................................................................................438
.................................................................................438
.....................................................................................438
BUS<b>:CPARallel:WIDTh
<BusWidth>
Sets the number of lines to be analyzed. D0 and D1 are reserved for chip select and
clock, all other digital channels can be used for data.
Suffix:
<b>
.
1..4
Selects the bus.
Note: SPI/SSPI and UART protocols occupy two bus lines.
Parameters:
<BusWidth>
Range:
1 to 14
Increment: 1
*RST:
4
Default unit: Bit
BUS<b>:CPARallel:CLOCK:SLOPe
<ClockSlope>
Selects if the data is sampled on the rising or falling slope of the clock, or on both edges
(EITHer). The clock slope marks the begin of a new bit
Suffix:
<b>
.
1..4
Selects the bus.
Note: SPI/SSPI and UART protocols occupy two bus lines.
Parameters:
<ClockSlope>
POSitive | NEGative | EITHer
BUS<b>:CPARallel:CS:POLarity
<Polarity>
Selects wether the chip select signal is high active (high = 1) or low active (low = 1).
Suffix:
<b>
.
1..4
Selects the bus.
Note: SPI/SSPI and UART protocols occupy two bus lines.
Command Reference