Measurements
R&S
®
RTM20xx
80
User Manual 1317.4726.02 ─ 01
Results: V1, V2,
Δ
V
Time
Sets two vertical cursor lines and measures the time from the trigger point to each cursor
point, the time between the two cursors and the frequency calculated from that time.
Results: t1, t2,
Δ
t, 1/t
Voltage & Time
Combines the "Voltage" cursor and "Time" cursor measurements. Two horizontal and
two vertical cursor lines are set and the voltages and time from the trigger point are
measured at the cursor positions, as well as the delta of the voltage and time values.
Results: t1, t2,
Δ
t, V1, V2,
Δ
V
Ratio X
Provides three cursors and measures the ratio of the x-values (e.g. a duty cycle) between
the first and second cursors and the first and third cursors:
(x2-x1)/(x3-x1)
The ratio is displayed as a floating value, in percent, in degrees and as a radian.
Results: abs, %, °,
π
Ratio Y
Provides three cursors and measures the ratio of the y-values (e.g. overshooting)
between the first and second cursors and the first and third cursors:
(y2-y1)/(y3-y1)
The ratio is displayed as a floating value and in percent.
Results: abs, %
Count
Provides three cursors to count signal transitions. The time base is defined by the first
two cursors, the third cursor defines the threshold value. As a result, the number of rising
and falling edges as well as the number of positive and negative impulses is counted.
Results:
,
, ,
Peak Values
Measures the positive and negative peak values between the two cursors, as well as the
absolute difference between the two peak values (peak-to-peak value):
Vpp = |(Vp+) - (Vp-)|
Results: Vp+, Vp-, Vpp
RMS, Mean,
σ
Measures the root mean square (RMS), the mean value, and the standard deviation of
measurement results between the two cursors.
Results: Mean, RMS,
σ
Cursor Measurements