Measurements
R&S
®
RTM20xx
91
User Manual 1317.4726.02 ─ 01
Delay
Measures the time difference on the middle reference level between two slopes of the
same or different waveforms. The measurement value is displayed as "Delay" in the result
table.
Select the sources and slopes in the
Set the middle reference level in the
Phase
Phase difference between two waveforms (time difference/period * 360) on the middle
reference level. The measurement value is displayed as "Phase" in the result table.
Select the waveforms in the
menu.
Set the middle reference level in the
Burst Width
Duration of one burst, measured from the first edge to the last edge that cross the middle
reference level. The measurement value is displayed as "Bst" in the result table.
Set the middle reference level in the
Count positive pulses
Count : Counts positive pulses within the displayed section of the waveform. A positive
pulse consists of a rising edge followed by a falling edge. The mean value of the signal
is determined. If the signal passes the mean value, an edge is counted. The pulse is
counted if a rising edge and a falling edge are detected. The number of positive pulses
is displayed as "Cnt
" in the result table.
Count negative pulses
Count
: Counts negative pulses within the displayed section of the waveform. A neg-
ative pulse consists of a falling edge followed by a rising edge. The mean value of the
signal is determined. If the signal passes the mean value, an edge is counted. The pulse
is counted if a falling edge and a rising edge are detected. The number of negative pulses
is displayed as "Cnt
" in the result table.
Count rising edges
Count : Counts transitions of the signal from low level to high level within the displayed
section of the waveform. To do so, the mean value of the signal is determined. If the
signal passes the mean value, a rising edge is counted. The number of rising edges is
displayed as "Cnt " in the result table.
Count falling edges
Count : Counts transitions of the signal from high level to low level within the displayed
section of the waveform. To do so, the mean value of the signal is determined. If the
signal passes the mean value, a falling edge is counted. The number of falling edges is
displayed as "Cnt " in the result table.
Trigger Freq.
Measures the frequency of the A-trigger signal based on the length of its period. The
measurement value is displayed as "f(Tr)" in the result table.
Automatic Measurements