Remote Control
R&S
®
RTM20xx
375
User Manual 1317.4726.02 ─ 01
BUS<b>:TYPE
<Type>
Defines the bus or interface type for analysis. For most types, a special option to the
instrument is required.
Suffix:
<b>
.
1..4
Selects the bus.
Note: SPI/SSPI and UART protocols occupy two bus lines.
Parameters:
<Type>
PARallel | CPARallel | I2C | SPI | SSPI | UART | CAN | LIN
*RST:
PARallel
BUS<b>:FORMat
<Format>
Sets the decoding format for the display on the screen.
Suffix:
<b>
.
1..4
Selects the bus.
Note: SPI/SSPI and UART protocols occupy two bus lines.
Parameters:
<Format>
ASCii | HEXadecimal | BINary | DECimal
*RST:
HEX
BUS<b>:DSIGnals
<BitsSignals>
Displays the individual bit lines above the decoded bus line.
Suffix:
<b>
.
1..4
Selects the bus.
Note: SPI/SSPI and UART protocols occupy two bus lines.
Parameters:
<BitsSignals>
ON | OFF
*RST:
ON
BUS<b>:DSIZe
<DisplaySize>
Sets the height of the decoded bus signal on the sreen.
Suffix:
<b>
.
1..4
Selects the bus.
Note: SPI/SSPI and UART protocols occupy two bus lines.
Parameters:
<DisplaySize>
SMALl | MEDium | LARGe
*RST:
MEDium