Remote Control
R&S
®
RTM20xx
294
User Manual 1317.4726.02 ─ 01
CHANnel<m>:DATA:HEADer?
Returns information on the channel waveform. For envelope waveforms, use the
command.
Table 15-12: Header data
Position
Meaning
Example
1
XStart in s
-9.477E-008 = - 94,77 ns
2
XStop in s
9.477E-008 = 94,77 ns
3
Record length of the waveform in Samples
200000
4
Number of values per sample interval, usually 1.
1
Suffix:
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Return values:
<DataHeader>
Comma-separated value list
Example:
-9.477E-008,9.477E-008,200000,1
Usage:
Query only
CHANnel<m>:DATA:ENVelope?
Returns the data of the envelope. The envelope consists of two waveforms. The wave-
forms data can be used in MATLAB, for example.
Use this command only for envelope waveforms. For other channel waveforms use
.
To set the range of samples to be returned, use
.
Suffix:
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Return values:
<Data>
List of values according to the format settings - the voltages of the
envelope points. The list contains two values for each sample
interval.
Usage:
Query only
CHANnel<m>:DATA:ENVelope:HEADer?
Returns information on the envelope waveform.
Use this command only for envelope waveforms. for all other channel waveforms use
.
Command Reference