Remote Control
R&S
®
RTM20xx
425
User Manual 1317.4726.02 ─ 01
is set to
ERRCondition
.
Parameters:
<SyncError>
ON | OFF
*RST:
OFF
TRIGger:A:LIN:ICONdition
<IdentifierCondition>
Sets the comparison condition for the identifier: If the pattern contains at least one X (don't
care), you can trigger on values equal or not equal to the specified value. If the pattern
contains only 0 and 1, you can also trigger on a range greater than or lower than the
specified value.
is set to
ID
or
IDDT
.
Parameters:
<IdentifierCondition> EQUal | NEQual | GTHan | LTHan
*RST:
EQ
TRIGger:A:LIN:IDENtifier
<Identifier>
Defines the identifier pattern.
is set to
ID
or
IDDT
.
Parameters:
<Identifier>
String containing binary pattern. Characters 0, 1, and X are
allowed. Enter the 6 bit identifier without parity bits, not the pro-
tected identifier.
TRIGger:A:LIN:DATA
<Data>
Defines the data pattern. The number of bytes in the data pattern is defined with
.
is set to
IDDT
.
Parameters:
<Data>
String containing binary pattern with max. 64 bit. Characters 0, 1,
and X are allowed. Make sure to enter complete bytes.
TRIGger:A:LIN:DCONdition
<DataCondition>
Sets the comparison condition for data: If the pattern contains at least one X (don't care),
you can trigger on values equal or not equal to the specified value. If the pattern contains
only 0 and 1, you can also trigger on a range greater than or lower than the specified
value.
is set to
IDDT
.
Command Reference