Remote Control
R&S
®
RTM20xx
267
User Manual 1317.4726.02 ─ 01
Bit No.
Meaning
6
User Request
This bit is set when the instrument is switched over to manual control.
7
Power On (supply voltage on)
This bit is set on switching on the instrument.
STATus:OPERation Register
In the
CONDition
part, this register contains information on which actions the instrument
is being executing. In the
EVENt
part, it contains information on which actions the instru-
ment has executed since the last reading. It can be read using the commands
STATus:OPERation:CONDition?
or
STATus:OPERation[:EVENt]?
.
See also:
The remote commands for the STATus:OPERation register are described in
ter 15.2.17.1, "STATus:OPERation Register"
Table 15-5: Bits in the STATus:OPERation register
Bit No.
Meaning
0
ALIGnment
This bit is set as long as the instrument is performing a self alignment.
1
SELFtest
This bit is set while the selftest is running.
2
AUToset
This bit is set while the instrument is performing an auto setup.
3
WTRIgger
This bit is set while the instrument is waiting for the trigger.
4 to 14
Not used
15
This bit is always 0.
STATus:QUEStionable Register
This register contains information about indefinite states which may occur if the unit is
operated without meeting the specifications. It can be read using the commands
Basics