Remote Control
R&S
®
RTM20xx
406
User Manual 1317.4726.02 ─ 01
..............................................................................................406
..........................................................................................411
CAN - Configuration
.........................................................................................406
.............................................................................................406
................................................................................407
BUS<b>:CAN:DATA:SOURce
<Source>
Sets the source of the data line. All channel waveforms can be used.
Suffix:
<b>
.
1..4
Selects the bus.
Parameters:
<Source>
CH1 | CH2 | CH3 | CH4 | D0..D15
*RST:
CH1
BUS<b>:CAN:TYPE
<SignalType>
Selects the CAN-High or CAN-Low line. CAN uses both lines for differential signal trans-
mission.
If you measure with a differential probe, connect the probe to both CAN-H and CAN-L
lines, and set the type CANH.
If you use a single-ended probe, connect the probe to either CAN_L or CAN_H, and select
the data type accordingly.
Suffix:
<b>
.
1..4
Selects the bus.
Parameters:
<SignalType>
CANH | CANL
*RST:
CANH
BUS<b>:CAN:SAMPlepoint
<SamplePoint>
Sets the position of the sample point within the bit in percent of the nominal bit time.
See also:
Suffix:
<b>
.
1..4
Selects the bus.
Command Reference