Display
R&S
®
RTM20xx
60
User Manual 1317.4726.02 ─ 01
Temper. Colors
Temperature Colors: Converts the brightness level of the displayed signals into a color
range, i.e. the color changes gradually in a wide color spectrum with increasing cumula-
tive occurance.
By default, rare values are displayed in blue, while more frequent values are red and very
frequent values are displayed in yellow or white, with various colors inbetween. This
results in a higher contrast, which provides a better perception of details in the waveforms.
Note:
The colors change if you invert the brightness display.
Remote command:
Aux. Cursors
Provides functions to display auxiliary cursors. Highlighted functions are active.
Trigger Event ← Aux. Cursors
Show or hides the point of the trigger. A small rhombus markes the intersection of the
trigger level and the trigger offset T.
The offset is related to the reference point and is displayed in the header of the display.
If you change the trigger level or the trigger position, a line appears temporarily to highlight
the changed value.
If the B-trigger is activated, the trigger is the result of an A-event followed by an B-event.
The trigger point has the color of the B-trigger source.
Channel Cursors ← Aux. Cursors
Activates or deactivates the channel cursor of the active channel. A channel cursor is a
line that marks the ground level of a channel. This line is displayed temporarily if the y-
scaling is modified and fades out automatically.
If a DC offset is defined, a second auxiliary line is displayed. The distance between those
two lines is the DC offset.
General Display Settings