Remote Control
R&S
®
RTM20xx
435
User Manual 1317.4726.02 ─ 01
15.2.14.1
Digital Channels
.................................................................................................435
...............................................................................................437
DIGital<m>:DISPlay
<State>
Enables and displays the indicated digital channel, or disables it.
Suffix:
<m>
.
0..15
Number of the digital channel
Parameters:
<State>
ON | OFF
*RST:
OFF
DIGital<m>:TECHnology
Selects the threshold voltage for various types of integrated circuits and applies it to the
channel group to which the indicated digital channel belongs.
Suffix:
<m>
.
0..15
Number of the digital channel.
Channel groups: 0..3; 4..7; 8..11; 12..15
Parameters:
<ThresholdMode>
TTL | ECL | CMOS | MANual
TTL: 1.4 V
ECL: -1.3 V
CMOS: 2.5 V
MANual: Set a user-defined threshold value with
*RST:
MAN
DIGital<m>:THReshold
Sets the logical threshold for the channel group to which the indicated digital channel
belongs.
Command Reference