FW 03.700
MEASurement<m>:STATistics:RESet
Deletes the statistical results for the selected measurement, and starts a new statistical
evaluation if the acquisition is running. The waveform count is set to 0 and all measure-
ment values are set to NAN.
Suffix:
<m>
.
1..4
Selects the measurement place.
Usage:
Event
Firmware/Software:
FW 03.700
MEASurement<m>:RESult:AVG?
<AverageValue>
Returns the average value of the current mesurement series.
The number of waveforms used for calculation is defined with
.
Suffix:
<m>
.
1..4
Selects the measurement place.
Query parameters:
<AverageValue>
Statistic value
Usage:
Query only
Firmware/Software:
FW 03.700
MEASurement<m>:RESult:STDDev?
<StandardDeviation>
Returns the statistical standard deviation of the current mesurement series.
The number of waveforms used for calculation is defined with
.
Suffix:
<m>
.
1..4
Selects the measurement place.
Query parameters:
<StandardDeviation> Statistic value
Usage:
Query only