Remote Control
R&S
®
RTM20xx
333
User Manual 1317.4726.02 ─ 01
Usage:
Query only
CURSor<m>:YDELta:SLOPe?
Returns the inverse value of the voltage difference - the reciprocal of the vertical distance
of two horizontal cursor lines:
1/
Δ
V
.
Suffix:
<m>
.
1
The numeric suffix is irrelevant.
Return values:
<DeltyYslope>
Inverse value
Usage:
Query only
CURSor<m>:YDELta[:VALue]?
Queries the delta of the values in y-direction at the two cursors.
Suffix:
<m>
.
1
The numeric suffix is irrelevant.
Return values:
<DeltaY>
Delta value in V
Usage:
Query only
CURsor<m>:XRATio:UNIT
<Unit>
Sets the unit for X Ratio measurements with
Suffix:
<m>
.
1
The numeric suffix is irrelevant.
Parameters:
<Unit>
RATio | PCT | GRD | PI
RATio - floating value
PCT - percent
GRD - degree
PI - radian
*RST:
RAT
CURSor<m>:XRATio[:VALue]?
Returns the ratio of the x-values (e.g. a duty cycle) between the first and second cursors
and the first and third cursors:
(x2-x1)/(x3-x1)
.
Set the unit of the result with
Command Reference