Display
R&S
®
RTM20xx
61
User Manual 1317.4726.02 ─ 01
Trigger Hyst. ← Aux. Cursors
Shows or hides the hysteresis range around the trigger level when the trigger level or the
hysteresis are changed. Hysteresis avoids unwanted trigger events caused by noise
oscillation around the trigger level.
See also:
Defaults ← Aux. Cursors
Restores the default cursor settings, i.e. both auxiliary cursors are displayed.
Grid
Defines how the grid is displayed.
"Lines"
Displays the grid as horizontal and vertical lines.
"Reticle"
Displays crosshairs instead of a grid.
"Off"
Removes the grid from the display.
Remote command:
VirtualScreen
Enables or disables the virtual screen.
If enabled, the virtual screen has 20 divisions, 8 of them are displayed. To select the
divisions to be displayed, press and turn the "Intensity / Virtual screen" knob at the left of
the screen.
If disabled, only 8 divisions are available, and all divisions are displayed.
Remote command:
4.2 XY-Diagram
XY-diagrams combine the voltage levels of two waveforms in one diagram. They use the
amplitude of a second waveform as the x-axis, rather then a time base. This allows you
to perform phase shift measurements, for example. With harmonically related signals the
resulting XY-diagrams are Lissajous patterns. XY-diagrams can also be used to display
the IQ representation of a signal.
It is also possible to define two source signals in y-direction for comparison.
The intensity of the XY-waveform can be set to a defined level, or be modulated dynam-
ically using a further source signal. In the latter case, the amplitudes of additional source
signal determine the intensity of the displayed waveform. For details, see
"Configuring the Intensity of an XY-Waveform"
XY-Diagram