Acquisition and Waveform Setup
R&S
®
RTM20xx
33
User Manual 1317.4726.02 ─ 01
Unit
Selects the unit that the probe can measure.
●
V - for voltage measurements
●
A - for current measurements
Remote command:
2.3.3.4
Probe Menu (Active Probe)
In the probe menu for active probes, you can also adjust the offset value for the selected
channel. Termination and attenuation are read out from the probe and do not need any
adjustment.
Access: CH N > "More > Probe"
Offset
See
ProbeMeter
Activates the integrated R&S ProbeMeter of active R&S probes and selects the probe
type. This voltmeter measures DC voltages between the probe tip and ground connection
with very high precision and enables ground-referenced measurements of voltages. The
DC measurement is performed continuously and in parallel to the measurements of the
oscilloscope. If activated, the DC offset measured on the probe tip is shown in a colored
label below the channel label.
"Inactive"
ProbeMeter is disabled.
"Single Ended" ProbeMeter of a single-ended active probe is enabled.
Remote command:
Micro Button
Active R&S probes have a configurable Micro Button on the probe head. Pressing this
button, you can perform an action on the instrument directly from the probe. During inter-
nal automatic processes the button is disabled, for example, during self alignment, auto-
set, and find level.
Select the action that you want to start from the probe.
Reference for Waveform Setup