Remote Control
R&S
®
RTM20xx
365
User Manual 1317.4726.02 ─ 01
15.2.12.5
Rise/Fall Time Search Configuration
.................................................................................365
........................................................................365
........................................................................365
................................................................................365
....................................................................................366
.................................................................................366
SEARch:TRIGger:RISetime:SLOPe
<Polarity>
Sets the slope to be found.
Parameters:
<Polarity>
POSitive | NEGative | EITHer
POSitive: to search for rise time.
NEGative to search for fall time.
EITHer: to search for rise and fall time
*RST:
POS
Firmware/Software:
FW 03.700
SEARch:TRIGger:LEVel:RISetime:LOWer
<LowerLevel>
Sets the lower voltage threshold. When the signal crosses this level, the rise time mea-
surement starts or stops depending on the selected slope.
Parameters:
<LowerLevel>
*RST:
400 mV
Default unit: V
Firmware/Software:
FW 03.700
SEARch:TRIGger:LEVel:RISetime:UPPer
<UpperLevel>
Sets the upper voltage threshold. When the signal crosses this level, the rise/fall time
measurement starts or stops depending on the selected slope.
Parameters:
<UpperLevel>
*RST:
600 mV
Default unit: V
Firmware/Software:
FW 03.700
SEARch:TRIGger:RISetime:RANGe
<Range>
Sets how the measured rise or fall time is compared with the given limit(s).
To set the rise/fall time, use
To set the range ± Δt, use
Command Reference