Triggers
R&S
®
RTM20xx
53
User Manual 1317.4726.02 ─ 01
Trigger On
Sets an additional delay condition for the B-event: time delay or event delay. According
to this selection, set also "Time" or "Events" to define the condition completely.
Remote command:
Time
Sets the time the instrument waits after an A-event until it recognizes B-events.
Remote command:
Events
Sets a number of B-trigger events that fulfill all B-trigger conditions but do not cause the
trigger. The oscilloscope triggers on the n-th event (the last of the specified number of
events).
Remote command:
Hysteresis A / B
Sets a hysteresis range around the trigger level. If the signal jitters inside this range and
crosses the trigger level thereby, no trigger event occurs. Thus, hysteresis avoids unwan-
ted trigger events caused by noise oscillation around the trigger level.
Hysteresis is available for the edge trigger (A-event and B-event) and for the width trigger.
To see the hysteresis on the display, activate "Trigger Hyst." in the "Display > Auxillary
Cursors" menu.
The hysteresis value depends on the vertical scale.
Remote command:
Reference for Triggers