Remote Control
R&S
®
RTM20xx
286
User Manual 1317.4726.02 ─ 01
ACQuire:WRATe
<WaveformRate>
Defines the mode to set the sample rate (samples per second saved in the memory) and
the waveform acquisition rate (waveforms per second).
Parameters:
<WaveformRate>
AUTO | MWAVeform | MSAMples | MANual
AUTO
To display the best waveform, the instrument selects the optimum
combination of waveform acquisition rate and sample rate using
the full memory depth (maximum record length).
MWAVeform
Maximum waveform rate: The instrument combines sample rate
and memory depth to acquire at maximum waveform acquisition
rate. In connection with persistence, the mode can display rare
signal anomalies.
MSAMples
Maximum sample rate: The instrument acquires the signal at max-
imum sample rate and uses the full memory depth. The result is a
waveform with maximum number of waveform samples, high
degree of accuracy, and low risk of aliasing.
MANual
The instrument acquires the signals at a sample rate that fills up
an user-defined record length. Set the record length using
ACQuire:POINts[:VALue].
*RST:
AUTO
CHANnel<m>:TYPE
<DecimationMode>
Selects the method to reduce the data stream of the ADC to a stream of waveform points
with lower sample rate.
Suffix:
<m>
.
The command affects all channels regardless of the indicated
channel number. The suffix can be omitted.
Parameters:
<DecimationMode>
SAMPle | PDETect | HRESolution
SAMPle
Input data is acquired with a sample rate which is aligned to the
time base (horizontal scale) and the record length.
PDETect
Peak Detect: the minimum and the maximum of n samples in a
sample interval are recorded as waveform points.
HRESolution
High resolution: The average of n sample points is recorded as
waveform point.
*RST:
SAMPle
Command Reference