4. Installation
4-4 Installing the main unit
33 / 191
4-4-2 Installing the main unit (using the mounting plate for compatibility with
the GD-70D)
CAUTION
Install so that there is no gap between the mounting plate for compatibility with the GD-70D and the
surface on which it is installed.
A gap may result in unnecessary vibrations and noise.
Before installing this product, remove the protective rubber caps fitted to GAS IN and GAS OUT.
If the product is installed with the rubber caps still attached, excess loads will be imposed on the
pumps and sensors when the power is turned ON. This may result in damage.
<ET models>
1 Use four double SEMS screws (M4/M5)
to secure the mounting plate for
compatibility with the GD-70D to the
installation surface.
You can also use three double SEMS
screws (M4/M5) when using the
mounting plate for compatibility with the
GD-70D for three or four mounting
screws.
The following screws are recommended
for mounting:
・
Screw type: double SEMS M4 or M5
・
Screw length: 8 mm or more
When securing with three double SEMS
screws, use the mounting plate for three
or four mounting screws. You cannot use
the mounting plate for four mounting
screws.
<Securing with four double SEMS screws>
Mounting plate for four
mounting screws
Mounting plate for three or four
mounting screws
<Securing with three double SEMS screws>
Mounting plate for three or
four mounting screws