7. Maintenance
7-5 Performing calibration
90 / 191
7-5-3 Connecting the calibration gas
With the pump stopped, connect the gas sampling bag with calibration gas and the gas sampling bag for
exhaust gas to the product.
<If calibration is to be performed for all sensors>
Connect the gas sampling bag with calibration gas to GAS IN on this product and connect the gas
sampling bag for exhaust gas to GAS OUT.
<If calibration is to be performed for each sensor>
Switch the chamber switching unit of this product to the flow passage for the sensor to be calibrated.
Connect the calibration gas.
1 Pull the projection on the top of the main unit
towards you to open the front cover.
The front cover opens to approximately 120°. Push
further down to open to 180°.
2 Loosen the black knurled screw in the center of
the chamber switching unit.
Gas
sampling
bag with
calibration
gas
Gas
sampling
bag for
exhaust
gas
Chamber
switching unit