7. Maintenance
7-3 Maintenance (regular maintenance) mode
81 / 191
7-3 Maintenance (regular maintenance) mode
Switch to maintenance mode to perform zero calibration, make span adjustments, and to make various
other settings.
7-3-1 Maintenance mode setting items
LCD display
Setting item
Details
2- 0 GAS TEST
Gas introduction test
Perform a gas introduction test.
2- 1 ZERO
Zero calibration
Perform zero calibration.
2- 2 SPAN
Span adjustment
Perform span adjustment.
2- 3 LAST CAL
Date last calibrated
Display the date on which calibration was last
performed.
2- 4 E VOLT
Set potential
Display the set potential.
2- 5 DEF FLOW
Flow value correction
Correct the flow value so that the flow becomes
0.6 L/min ± 10 % (0.54 - 0.66).
2- 6 FLOW
Flow display, flow
adjustment
Adjust the pump and display the current flow.
2- 7 TEMP
Ambient temperatures
Display the current temperature in the installation
environment.
2- 8 ----
Not used
2- 9 SETTING1
Environmental setting 1
Make settings for INHIBIT, alarm setpoints, etc.
Perform fault tests and maintenance. (Refer to
‘7-9 Making changes in Environmental setting 1’.)
2-10 SETTING2
Environmental setting 2
Make settings for items such as alarm patterns, fault
patterns, zero suppression values, date and time, and
external output. (Refer to ‘7-10 Making changes in
Environmental setting 2’.)
2-11 ----
Not used
2-12 FAULT
Not used
2-13 S SET
Sensor start setting
Set the date of first use for the sensor.
2-14 F MODE
To factory mode
Not used
2-15 RETURN
To user mode
Switches to user mode.