5. Usage Instructions
5-6 Checking setting values (user mode)
59 / 191
NOTE
The first alarm setpoint, second alarm setpoint, and alarm delay time can be set in Environmental
setting 1 in maintenance mode. (Refer to ‘7-9 Making changes in Environmental setting 1’.)
The zero suppression value, zero follower ON/OFF setting, and sensitivity correction ON/OFF
setting can be set in Environmental setting 2 in maintenance mode. (Refer to ‘7-10 Making changes
in Environmental setting 2’.)
5-6-2 Checking flow values (FLOW)
Display the current flow value (L/min).
1 Press
the
▲
key or
▼
key in user mode to select
[1- 3 FLOW], then press the TEST/SET key.
2 Check the current flow value (L/min).
Displays the total value for flow passage A and flow passage B.
3 Press the MODE key.
The display reverts to the [1- 3 FLOW] display.