4. Installation
4-4 Installing the main unit
30 / 191
4-4-1 Installing the main unit (without the mounting plate for compatibility
with the GD-70D)
CAUTION
Before installing this product, remove the protective rubber caps fitted to GAS IN and GAS OUT.
If the product is installed with the rubber caps still attached, excess loads will be imposed on the
pumps and sensors when the power is turned ON. This may result in damage.
<ET models>
1 Place the main unit against the installation surface and
secure with two screws.
We recommend double SEMS screws (M4-14).
2 Pull the projection on the top of the main unit
towards you to open the front cover.
The front cover opens to approximately 120°. Push
further down to open to 180°.
3 Use two screws to secure the main unit to the
installation surface.
Projection on
the top
Содержание GD-84D-EX Series