11. Product Specifications
11-2 Sensor specifications
177 / 191
11-2 Sensor specifications
11-2-1 Electrochemical type
Sensor model
ESF
Detection target gas
Toxic gases
*1
Oxygen
Detection range
Depends on detection target gas.
0 -25 vol%
Detection method
Pump suction type
Alarm setpoints
Depends on detection target gas.
18 vol%(default)
Alarm accuracy
(for identical
conditions)
Within ±30 %
*2
Within ±0.6 vol%
Alarm delay time
(for identical
conditions)
Within 60 seconds
*3
Within 10 seconds
*4
Alarm type
Two-stage alarm (H-HH)
Two-stage alarm (L-LL, L-H)
Operating
temperature range
-10 – 40 °C (no sudden changes)
Operating humidity
range
20 – 90 % RH (No condensation) (Specifications differ for certain gases.)
*1 The range of operating temperatures and operating humidity may differ depending on the detection target gas.
*2 Indicates the precision against the alarm setpoint.
*3 When gas is provided at 1.6 times the alarm setpoint. (excluding piping and communication delays)
*4
This is the time required to reach the 18 vol% alarm point after supplying gas with a concentration of 10 to 11 vol%
(excluding piping and communication delays)
NOTE
Contact Riken Keiki for specifications and information on other sensors.
The number of years of ESF sensor warranty depends on the sensor model. Please refer to the
sensor specification sheet.