7. Maintenance
7-3 Maintenance (regular maintenance) mode
83 / 191
Switches to maintenance mode.
After making the settings, hold down the MODE key for
about three seconds in maintenance mode to switch to
detection mode.
WARNING
Once you are done, hold down the MODE key in maintenance mode to return to detection mode.
The product will not detect gases while in the adjustment/setting state. The product will revert to
detection mode in 10 hours if left in maintenance mode. Exception: Due to the possibility that
replacement tasks are underway, the device will not revert to detection mode if left in maintenance
mode while setting environmental setting 1 is underway ([2- 9 SETTING1] >[SET-3 MAINTE]). This
is dangerous because no monitoring occurs during this time. Keep this in mind.
NOTE
If password setting is ON, a password must be entered to switch to maintenance mode.
The password is set beforehand by Riken Keiki. Please contact our sales department if you do not
know the password.
Select [2-15 RETURN] in maintenance mode and press the TEST/SET key to switch to user mode.