7. Maintenance
7-10 Making changes in Environmental setting 2
113 / 191
7-10-4 Settings for contacts during alarm tests (TEST RLY)
Set the operation of contacts during alarm tests ON or OFF.
Set to ON to allow contacts to operate during alarm tests.
The default setting is OFF.
1 In maintenance mode, press the
▲
key or
▼
key to
select [2-10 SETTING2], then press the TEST/SET
key.
2 Press
the
▲
key or
▼
key to select [SET- 4 TEST RLY],
then press the TEST/SET key.
3 Press
the
▲
key or
▼
key to select [OFF] or [ON],
then press the TEST/SET key.
The display reverts to the [SET- 4 TEST RLY] display.
4 Press the MODE key.
The display reverts to the [2-10 SETTING2] display.
NOTE
Press the MODE key in step 3 to cancel the settings and return to the [SET- 4 TEST RLY] display.