8. Replacing and cleaning parts
8-2 Replacing periodic replacement parts
154 / 191
12 Return the chamber unit to its original position.
13 Tighten the four white knurled screws on the
chamber unit.
Tighten the knurled screws so that the chamber
unit is secured in place.
Gas leaks may result if the knurled screws are not
tightened properly.
14 Close the front cover on the main unit.
Close the front cover firmly so that it clicks back into place.
15 Once replacement is complete, press the MODE key.
The display returns to the [SET- 3 MAINTE] display. The pump and fan will restart.
16 Press the MODE key.
The display reverts to the [2- 9 SETTING1] display.
Chamber unit