7. Maintenance
7-10 Making changes in Environmental setting 2
120 / 191
7-10-8 Alarm pattern setting (ALM PTRN)
Set the alarm patterns for the alarm setpoints.
Set to either auto-reset ([NL]) or self-latching ([L]).
The default setting is auto-reset ([NL]).
NOTE
The alarm pattern settings will affect how the device behaves. For normal use, leave this setting
unchanged.
1 In maintenance mode, press the
▲
key or
▼
key to
select [2-10 SETTING2], then press the TEST/SET key.
2 Press
the
▲
key or
▼
key to select [SET- 8 ALM PTRN],
then press the TEST/SET key.
3 Press
the
▲
key or
▼
key to select [NL] or [L],
then press the TEST/SET key.
The display reverts to the [SET- 8 ALM PTRN] display.
4 Press the MODE key.
The display reverts to the [2-10 SETTING2] display.
NOTE
Press the MODE key in step 3 to cancel the settings and return to the [SET- 8 ALM PTRN] display.