8. Replacing and cleaning parts
8-2 Replacing periodic replacement parts
152 / 191
8-2-3 Flow sensor unit replacement
If the main unit is running, stop the pump and fan before replacing the flow sensor unit.
1 In maintenance mode, press the
▲
key or
▼
key to
select [2- 9 SETTING1], then press the TEST/SET key.
2 Press
the
▲
key or
▼
key to select [SET- 3 MAINTE],
then press the TEST/SET key.
The pump and fan stop.
3 Pull the projection on the top of the main unit
towards you to open the front cover.
The front cover opens to approximately 120°. Push
further to open to 180°.
4 Loosen the four white knurled screws on the
chamber unit.
5 Grasp the projections on the left and right of
the chamber unit. Pull towards you and lower
to open.
Chamber unit
projection
Chamber unit
projection