8. Replacing and cleaning parts
8-1 Replacing consumable parts
145 / 191
3 Pull the projection on the top of the main unit
towards you to open the front cover.
The front cover opens to approximately 120°. Push
further to open to 180°.
4 Loosen the four white knurled screws on the
chamber unit.
5 Grasp the projections on the left and right of
the chamber unit. Pull towards you and lower
to open.
6 Press
the
▲
key or
▼
key to select the sensor
to replace, then press the TEST/SET key.
Multiple sensors can be selected.
Pressing the
▼
key with the [B2] sensor selected
will initiate suction of all gases. Do not press the
▼
key while the [B2] sensor is selected until you have
returned the chamber unit to its original position.
The following information is displayed, depending
on the status of the selected sensor:
・
[Not any]:
Not connected
・
[Dummy]:
Dummy connection
・
[C-01][C-02]: The sensor was replaced.
・
[C-03][C-04]: Problem with sensor combination or
the sensor is incompatible.
・
[USED]:
The sensor may be older than the one
previously fitted.
・
[COMB]:
Problem with sensor combination
Chamber unit
projection
Chamber unit
projection