7. Maintenance
7-5 Performing calibration
89 / 191
10
Disconnect the calibration gas from this product.
When span adjustment is complete, disconnect the span gas sampling bag and the gas sampling bag
for exhaust gas. (Refer to ‘7-5-4 Disconnecting the calibration gas’.)
Return the chamber switching unit to the [D] (detection mode) position.
11 Press the MODE key.
The display reverts to the [2- 2 SPAN] display.
12 Hold down the MODE key for about three seconds to switch to detection mode.
NOTE
Press the SHIFT key in step 2 to display the span adjustment margin for the selected sensor as you
hold down the key. Restart to reset the displayed margin.