Two Channel Gas Detector

E2632

User Manual

Содержание E2632

Страница 1: ...Two Channel Gas Detector E2632 User Manual...

Страница 2: ...nts 5 Operating conditions 5 Installation guidelines 6 Mounting dimensions 7 Connections 8 Operation 9 Sensor probe handling 10 The electrochemical gas sensor replacement procedures 10 Maintenance 11...

Страница 3: ...ioxide SO2 0 10 ppm 0 100 ppm Electrochemical 2 years 30 sec 6 months Ammonia NH3 0 100 ppm 0 500 ppm Electrochemical 2 years 50 sec 6 months Methane CH4 0 100 LEL Metal oxide semiconductor 5 years 60...

Страница 4: ...max 5 A 30 VDC 250 VAC Alarm signaling Blinking LEDs buzzers 85 dB LEDs Green red operation fault red gas alarm Enclosure Grey ABS wall mount protection class IP65 Dimensions H123 W201 D55 mm Output c...

Страница 5: ...components directly as they are sensitive to static electricity Connection diagrams can be found in the connections section The device might not perform correctly or be damaged if the wrong power sup...

Страница 6: ...e area of the device However the coverage area for a detector is usually between 2 5 to 5 meters radius For early leakage detection install the sensor as close as possible to the potential leakage sou...

Страница 7: ...E2632 Rev 17 05 2021 Mounting dimensions 7...

Страница 8: ...of the external devices to each sensor Connect the power terminals N and L to the 24 V AC DC source if you are using detector version 24 or to 230 V AC mains if you are using detector version 230 see...

Страница 9: ...the sensor to stabilize For stable operating it is recommended to keep the detector powered constantly except for periods of maintenance and calibration replacement etc Operation During the first ca 6...

Страница 10: ...l launch blinking and beeping stops as soon as the key is withdrawn Sensor probe handling The sensor probes of all types are equipped with a hydrophobic microporous PTFE filter to protect the sensor f...

Страница 11: ...fore delivery Provided that the sensor is used under moderate conditions field recalibration is recommended every few months The calibration intervals are stated in the gases table Please contact your...

Страница 12: ...nufacturer will at its option either repair or replace a product that proves to be defective This warranty is void if the product has been operated in conditions outside ranges specified by the Manufa...